Not a valid Time
This is a required field.

Nuvair PRO DUO O2/CO2 Multigas Alarm Analyzer

SKU
NUV-9618-LB
$2,292.00
PRO DUO O2/CO2 PANEL MOUNT ANALYZER W/LITHIUM BATTERY IN WATERTIGHT BOX

Restrictions / Notifications

  • Hazardous Material
    This item contains hazardous materials and additional freight cost and restrictions may apply.
  • Non-Returnable/Non-Cancelable
    This item cannot be canceled or returned once an order has been placed.
  • CA PROP65
    Warning: Cancer and Reproductive Harm www.p65warnings.ca.gov
  • MSDS
    Safety Data Sheet available online or by request.
More Information
Features
  • Monitors oxygen and carbon dioxide levels to ensure compliance with breathing air standards for CGA, NFPA, OSHA, CSA, EN norms, and others
  • Oxygen reading in any mix in the range 0.0-100.0% of volume
  • Carbon dioxide reading in parts per million (ppm)
  • Two (2) custom audible and visual alarms for oxygen (O₂) & carbon dioxide (CO₂)
  • Robust waterproof hard case
  • Automatic calibration of O2 and CO2 sensors
  • Temperature-compensated sensors
  • User-replaceable battery and sensors
  • Low battery warning indicator (battery-powered versions)
  • Factory reset
  • Fast response
  • Compact water-resistant container
  • Designed to test breathing gases
  • Easy to operate
  • Reliable and accurate
  • Optional relays for external alarm or compressor control
  • Optional flow meter
  • Audible visual alarms for high moisture content in filters
  • Warning light for filter changes

 

Sensor Specifications
Sensor Pro O2 Pro CO2
Flow Rate 0.5-2 L/min 0.6-1 L/min
Resolution ±0.1% 50 ppm resolution from 0 to 1000 ppm, then 100 ppm up to full scale
Repeatability N/A ±2% of full scale @ 68°F (20°C) ambient
Linearity N/A At ambient temperature and pressure: ±2% FSD or ±10% of the reading, whichever is greater
Accuracy Within ±1% of full scale at constant temperature and pressure (0-1 atm) when calibrated with 100% oxygen ±2% over full scale
Sensor Type Electrochemical Non-dispersive infrared (NDIR)
Expected Sensor Life Up to 36 months in air at 77°F (25°C) and 50% rh 5-10 years based on analyzer handling and use. Check sensor accuracy frequently with test gas and attempt to recalibrate if reading is inaccurate. Replace sensor if recalibration fails.
Range 0-100% oxygen (max), 0-1% oxygen (min) 0-2000 ppm

Sensor Pro O2 Pro CO2
Flow Rate 0.5-2 L/min 0.6-1 L/min
Resolution ±0.1% 50 ppm resolution from 0 to 1000 ppm, then 100 ppm up to full scale
Repeatability N/A ±2% of full scale @ 68°F (20°C) ambient
Linearity N/A At ambient temperature and pressure: ±2% FSD or ±10% of the reading, whichever is greater
Accuracy Within ±1% of full scale at constant temperature and pressure (0-1 atm) when calibrated with 100% oxygen ±2% over full scale
Sensor Type Electrochemical Non-dispersive infrared (NDIR)
Expected Sensor Life Up to 36 months in air at 77°F (25°C) and 50% rh 5-10 years based on analyzer handling and use. Check sensor accuracy frequently with test gas and attempt to recalibrate if reading is inaccurate. Replace sensor if recalibration fails.
Range 0-100% oxygen (max), 0-1% oxygen (min) 0-2000 ppm
Analyzer Specifications
  • Alarms: Two user-programmable audible and visual alarms
  • Power: Rechargeable lithium battery with 110/230V charger
  • Operating Temperature: 32-104°F (0-40 °C)
  • Storage Temperature: 32-122°F (0-50 °C)
  • Operating Humidity: 0-95% rh (non-condensing)
  • Power: Rechargeable lithium battery or 110-220 V wall plug-in
  • Dimensions (L x W x H): 3.6 x 8.8 x 7.5 in. (9.3 x 22.5 x 19 cm)
  • Weight: 4 lbs. (0.95 kg)

 

*All analyzer specifications are at ambient/sea level, 77°F (25 °C) and are subject to change without notice.

 

WARNING: Never expose gas sensors to pressure, or you may cause damage and/or false readings. Damaged sensors will not provide accurate gas analysis. Most gas analyzers can be used to analyze a regulated gas sample flow, the contents of a gas cylinder, or the flow from a regulator. The flow rate of gas must equal 1-5 L/min. To produce this flow, a Flow Restrictor and Regulator may be required. A faulty Flow Restrictor can lead to a false analyzer reading. Flow Restrictors should be regularly tested with a Flow Meter. Inaccurate gas analysis can lead to serious personal injury or death.

The Nuvair PRO DUO series of gas analyzers come in different pairings to match your multi-gas analyzing requirements. Mounted in a rugged, water-resistant carrying case, the PRO DUO has two user-programmable audible and visual alarms for each gas sensor.

 

The Nuvair Pro Duo series of multi-gas alarm analyzers are mounted in rugged, water-resistant cases. Each sensor has two user-programmable audible and visual alarms. The internal sensors are plumbed together to monitor a single input. The analyzer has visual and optional audible alarms for each gas.

 

An audible alarm sounds if any alarm set point is reached. When in alarm, the analyzer's analog output can activate an optional relay that shuts down compressors automatically, sounds a remote alarm, or activates emergency backup systems. All -LB models come equipped with a rechargeable lithium-ion battery and include a universal 110/230-volt charger.

The Nuvair Pro O2/CO2 multigas alarm analyzer is mounted in a rugged, water-resistant case. Each sensor has two user programmable audible and visual alarms. The internal sensors are plumbed together to monitor a single input. The analyzer has visual and optional audible alarms for each gas. An audible alarm sounds if any alarm set point is reached. When in alarm, the analyzer's analog output can activate an optional relay that shuts down compressors automatically, sounds a remote alarm, or activates emergency backup systems. Comes equipped with a rechargeable lithium-ion battery and universal 110/230-volt charger.

Amron International Customer Service
We respond to most inquiries within 24 hours, Monday – Friday, excluding holidays.
Phone
Toll-Free: (877) 462-6700
Local & International: (760) 208-6500
Fax: (760) 599-3857
Email
sales@amronintl.com
Hours of Operation
Customer Service
Monday - Friday, 8am – 5pm Pacific Time Zone

Warehouse/Will Call Orders
Monday – Friday, 9am – 3:30pm Pacific Time Zone
We found other products you might like!
  1. Nuvair 9611 Pro Oxygen (O2) Alarm Analyzer Handheld Series
    Selection Tool
    Nuvair Pro Oxygen (O2) Alarm Analyzer Handheld Series
    SKU: 9611 Series
    SHIP NOW
    As low as $590.00