Replacement carbon dioxide (CO2) sensor for the following Nuvair Pro CO2 analyzers:
9616
9617
Specifications:
Flow Rate: 0.6-1 L/min
Resolution: 50 ppm resolution from 0 to 1000 ppm, then 100 ppm up to full scale.
Repeatability: ± 2% of full scale @ 68°F (20°C) ambient.
Linearity: At ambient temperature and pressure: ± 2% FSD or ± 10% of the reading, whichever is greater.
Sensor Type: NDIR (Non-dispersive infrared)
Expected Sensor Life: 5-10 years based on analyzer handling and use.
Replacement Procedure: Sensor can be replaced by the user.
Range: 0-2,000 ppm
Operating & Storage Temperature: -4° to +122°F (-20° to +50°C)
Operating Humidity: 0 to 95% rh, non-condensing
Weight: 0.2 oz. (6 g)
* All specifications are at ambient/sea level, 77°F (25°C) and are subject to change without notice. Calibration must be confirmed with a calibration CO2 test gas.
*WARNING: Never expose gas sensors to pressure or you may cause damage and/or false readings. Damaged sensors will not provide accurate gas analysis. Most gas analyzers can be used to analyze a regulated gas sample flow, the contents of a gas cylinder, or the flow from a regulator. The flow rate of gas must equal 1-5 L/min. To produce this flow, a Flow Restrictor and Regulator may be required. A faulty Flow Restrictor can lead to a false analyzer reading. Flow Restrictors should be regularly tested with a Flow Meter. Inaccurate gas analysis can lead to serious personal injury or death..
Nuvair's Carbon Dioxide (CO2) Sensor Replacement for Pro Carbon Dioxide (CO2) Analyzers NUV-9616-SEN is a replacement for the following Nuvair Pro CO2 analyzers: 9603, 9604-LB 9615-LB, 9615.5-LB, 9616, 9617, 9635, and 9641-NV.