Application Range:

  • Standard Measuring Range: 1 to 5 ppm
  • Number of Strokes n: 6
  • Time for Measurement: approx. 3 min
  • Standard Deviation: ± 30 %
  • Color Change: light blue → light pink

Ambient Operating Conditions:

  • Temperature: 0 to 40 °C
  • Absolute Humidity: 15 to 90 % r.h.
  • At 0 to 10 °C, concentrations of sulfuryl fluoride are displayed with approximately half sensitivity.
  • At 30 to 40 °C and air humidity < 30 % r.h. the displays can only be recognised as of > 2 ppm.
  • At 30 to 40 °C and air humidity > 75 % r. h., concentrations of sulfuryl fluoride are displayed with approximately half sensitivity.

Reaction Principle:

  • a) pyrolysis sulfuryl fluoride → HF
  • b) HF + zircon / quinalizarin → pink reaction product (HF destroys the quinalizarin/zircon complex by complexation of the zircon)

Cross Sensitivity:

  • Fluorinated hydrocarbons are also displayed with different sensitivities.
  • Ammonia and other basic gases could, depending on the concentrations, either shorten or prevent the color change.
  • The following chemicals have no influence on the display of 3 ppm sulfuryl fluoride:
    • 2 ppm formaldehyde, 5 ppm methyl bromide and 1 ppm phosphine.
    • When the oxygen concentration decreases, the sensitivity decreases.
      • For example, the 3 ppm display at 18 % oxygen is very weak.

Additional Information:

  • Do not use in areas subject to explosion hazards.
  • Tubes get hot.
  • During and shortly after measurement, do not touch the first tube near the black pyrolysis layer.
Draeger Sulfuryl Fluoride 1/a Detector Tubes (5 tests per box) 8103471
ATTACHMENTS
Amron International Customer Service
We respond to most inquiries within 24 hours, Monday – Friday, excluding holidays.
Phone
Toll-Free: (877) 462-6700
Local & International: (760) 208-6500
Fax: (760) 599-3857
Email
sales@amronintl.com
Hours of Operation
Customer Service
Monday - Friday, 8am – 5pm Pacific Time Zone

Warehouse/Will Call Orders
Monday – Friday, 9am – 3:30pm Pacific Time Zone
We found other products you might like!