Application Range:

  • Standard Measuring Range: 100 to 2,500 ppm for n-octane.
  • Number of Strokes n: 2
  • Time for Measurement: approx. 30 s
  • Standard Deviation: ± 10 to 15 %
  • Color Change: white → green

Ambient Operating Conditions:

  • Temperature: 0 to 40 °C
  • Absolute Humidity: < 30 mg H2O / L

Reaction Principle:

  • C8H18 + I2O5 → I2

Cross Sensitivity:

  • Many other petroleum hydrocarbons are indicated as well, but with different sensitivities.
  • It is impossible to differentiate between them.
  • Aromatics are indicated, but only with low sensitivities.
  • Carbon monoxide is indicated in comparable concentrations with about half the sensitivity.