Draeger Hydrocyanic Acid 0.5a Detector Tubes (10 tests per box) 8103601

Restrictions / Notifications

  • CA PROP65
    Warning: Cancer and Reproductive Harm www.p65warnings.ca.gov
  • MSDS
    Safety Data Sheet available online or by request.
More Information

Application Range:

  • Standard Measuring Range: 0.5 to 5 ppm / 5 to 50 ppm
  • Number of Strokes n: 10 / 2
  • Time for Measurement: Approx. 2.5 min / CA. 0.5 min
  • Standard Deviation: ± 10 to 15 %
  • Color Change: Yellow → Red

Ambient Operating Conditions:

  • Temperature: 0 to 40 °C
  • Absolute Humidity: < 40 mg H2O / L

Reaction Principle:

  • a) HCN + HgCl2 ä HCl
  • b) HCl + methyl red → red reaction product

Cross Sensitivity:

  • 30 ppm hydrogen sulfide, 300 ppm ammonia, 40 ppm sulfur dioxide, 20 ppm nitrogen dioxide and 1,000 ppm hydrogen chloride do not affect the indication.
  • Hydrogen sulfide causes the pre-layer to discolor to dark brown.
  • Ammonia concentrations above 300 ppm can cause the indication at the beginning of the indicating layer to discolor back to yellow.
  • Acylonitrile up to a concentration of 1,000 ppm does not affect the indication.
  • It is impossible to measure hydrocyanic acid in the presence of phosphine.
Draeger Hydrocyanic Acid 0.5a Detector Tubes (10 tests per box) 8103601
Amron International Customer Service
We respond to most inquiries within 24 hours, Monday – Friday, excluding holidays.
Phone
Toll-Free: (877) 462-6700
Local & International: (760) 208-6500
Fax: (760) 599-3857
Email
sales@amronintl.com
Hours of Operation
Customer Service
Monday - Friday, 8am – 5pm Pacific Time Zone

Warehouse/Will Call Orders
Monday – Friday, 9am – 3:30pm Pacific Time Zone
We found other products you might like!