Amron International 8891-07 Oxygen Treatment Hood Neck Ring Assembly with Multipurpose Plug and O-Ring

SKU: 8891-07 Rate this product

$53.50

Availability: 15 In Stock

Click to Enlarge Image

Overview

Features:

  • Silicone Neck Seal 8891-009-00 is NOT compatible with OLD Neck Ring P/N: 8891-001
  • Silicone Neck Seal 8891-009-00 is ONLY compatible with 2014 improved Neck Ring P/N: 8891-011
  • Latex Neck Seals (Trimmed P/N: 8891-005 and Untrimmed P/N: 8891-005-00) are compatible with BOTH OLD Neck Ring (P/N: 8891-001) and 2014 improved Neck Ring (P/N: 8891-011)
  • Both versions of the Oxygen Treatment Hood P/N: 8891-010 fit BOTH OLD and 2014 improved Neck Rings
  • CE and FDA Registered
  • Patent Pending
  • Neck Ring Material: Polycarbonate (Autoclavable)
  • For use with: Amron O2 Vinyl Hood with Optical Window
  • 2014 improved design: oblong shape
  • 2014 improved design: larger opening makes it easier to place over patient’s head
  • 2014 improved design: two square neck seal grooves
  • Color: white
  • Replaces P/N: 8891-001
  • Compatible with Models 8891, 8892, 8893 & 8891-010

Sterilization:

Please reference the User Manual located in the Attachments Tab on this web page for recommended cleaning instructions. Amron recommends SaniZide Plus as a convenient, fast-acting germicidal disinfectant. Gas sterilization can be used if close attention is paid to temperature, but multiple cycles of gas sterilization will shorten the working life of the product. Gas sterilization temperatures must be kept below 130 degrees F to avoid affecting the shape of the hood ring. See the FAQ tab on this web page for additional information about gas sterilization.

Oxygen Treatment Hood Neck Ring Assembly - P/N: 8891-07 Includes:

  • P/N 8891-011 Neck Ring
  • P/N 8891-25 Multipurpose Plug
  • P/N 8891-15 O-ring

Caution: Federal law (USA) mandates that Amron can only sell this product to physicians and/or licensed clinical practitioners.

Description

Amron International’s Oxygen Treatment Hood Neck Ring Assembly with Multipurpose Plug and O-Ring is for use with Amron’s Oxygen Treatment Hood with Optical Viewing Window. This product has been recently redesigned based directly on your feedback and is CE and FDA Registered.

The oblong shaped neck ring is larger than its predecessor, making it much easier for the attendant to place the Amron hoods over the patient’s head (essentially the larger neck ring now has a universal fit where one size fits all patients). The redesigned neck ring now has two square neck seal grooves on it, which provides for a more secure attachment of the neck seal to the neck ring.

The neck ring was developed to withstand multiple cycles in an autoclave, gas sterilized or cleaned using SaniZide disinfectant provided by Amron. The reusable neck ring has a conveniently located multipurpose plug port as well as supply and exhaust hose ports, with an accompanying O-ring seal for the vinyl hood attachment (Note: The 8891-07 Neck Ring, 8891-15 O-ring, and 8891-25 multipurpose plug come installed together as a set on the 8891-07 Neck Ring Assembly). The supply and exhaust locations allow oxygen to circulate easily through the hood at reduced noise levels. As a soft flow of gas helps to cool the patient, the optical-quality viewing window is continuously being defogged.

Caution: Federal law (USA) mandates that Amron can only sell this product to physicians and/or licensed clinical practitioners.

Note: Patent Pending

Attachments
Videos

Additional Information

Brand Amron International
Video No
FAQ
 

Product FAQ's

Ask a Question

  • The method of cleaning as stated in the general instructions does not suit our unit due to the clean area that would be required to drip dry the hoods. Is there any other approved method of cleaning?

    We only have two methods approved for cleaning.

    1)     The first is spraying Sanizide and letting the hood sit for 10 minutes.

    2)    The second is gas sterilization using the method below:

    1. Low - 4  hours at 100 degrees Fahrenheit with an aeration time of 36 hours
    2. High - 1 hour at 130 degrees Fahrenheit with an aeration time of 12 hours

    Using the gas sterilization will shorten the life of the hood but we have many customers that use it and are very happy with the results.

     